These models have different corresponding design standards, and it doesn't matter if it's good or not! CZF-3 corresponds to GB2408-1996 "Test Method for Burning Performance of Plastics - Horizontal Vertical and Vertical Method"; CZF-5 corresponds to GB2408-2008 "Test Method for Burning Performance of Plastics Horizontal Vertical and Vertical Method", (its content is equivalent to the standard of UL94); CZF-6 corresponds to the GB T12960-1996 "Quantitative Determination of Cement Components" standard, and HVR-JT is the setting of some manufacturers for the model, and its corresponding design standard is consistent with UL94; JXW-056AB is the same as GB T8333-2008 "Test Method for Combustion Performance of Rigid Foam Plastic - Vertical Combustion Method". The relevant standards involved are in the technical information section of the official website of Nanjing Jionglei Instrument, and it is recommended that you can choose to buy the standard according to your actual needs.
-
Selection of hipot tester.
The most important thing to choose a withstand voltage meter is two indicators, the maximum output voltage value and the maximum alarm current value must be greater than the voltage value and alarm current value you need. In general, the standard of the tested product stipulates the value of the applied high voltage and the alarm judgment current value. If the higher the applied voltage, the larger the alarm judgment current, then the greater the power of the step-up transformer that needs to be built-up, and the power of the step-up transformer of the general withstand voltage meter is etc.
The maximum voltage can reach tens of thousands of volts. The maximum alarm current is 500mA-1000mA and so on. Therefore, when choosing a pressure tester, you must pay attention to these 2 indicators.
If the power is too large, it will cause waste, and if the power is too small, the withstand voltage test cannot correctly judge whether it is qualified or not. According to IEC414 or (GB6738-86) specified in the selection of the power method of the withstand voltage meter, we believe that it is more scientific. "First, the output voltage of the hipot meter is adjusted to 50% of the specified value, and then the test product is connected, and when the observed voltage drop is less than 10% of the voltage value, the power of the hipot meter is considered to be sufficient.
That is, if the voltage value of the withstand voltage test of a product is 3000 volts, first adjust the output voltage of the withstand voltage meter to 1500 volts and then connect the tested product, if the output voltage of the withstand voltage meter drops less than 150 volts at this time, then the power of the withstand voltage meter is sufficient. There is a distributed capacitance between the live part of the test product and the housing. The capacitor has a cx capacitive reactance, and when an alternating current is applied to the two ends of the cx capacitor, a current is induced.
The magnitude of this current is proportional to the capacity of the CX capacitor and the applied voltage, when the current is large enough or exceeds the maximum output current of the withstand voltage meter, the withstand voltage meter cannot correctly determine whether the test is qualified or not.
Huaneng Solar Electric is specialized in production.
Microcomputer relay protection tester.
High Voltage Test Recruitment.
Earth Resistance Tester.
SAW resonators.
Volumetric meter. ST403A High Voltage Generator.
Microcomputer relay protection tester.
How resolvers work.
Subwoofer. Outdoor high-voltage load switch.
Generator rotor AC impedance tester.
Transformer Loss Tester.
If you need to understand other parameters of the above equipment, please take a look at "Wuhan Huaneng Sunshine Electric".
-
Huaneng Solar Electric is specialized in production.
Empty load. Underground cable detectors.
Lightning protection components.
How to use the resistance meter.
High Voltage Switching Characteristics Tester.
Insulation resistance meter verification device.
Resonant change. What is relay protection tuning.
High-voltage filter capacitors and discharge rods.
SAW resonators.
Serial resonance test device.
35kV cable withstand voltage test.
If you need to understand other parameters of the above equipment, please take a look at "Wuhan Huaneng Sunshine Electric".
-
1. Turn on, the first interface has three test methods: horizontal method, vertical method and self, and the selection key of the point response enters the corresponding interface.
2. ta - the following value is the flame application time, you can directly click the number to enter the keyboard interface to modify the time.
3. The functions of the common keys of start, stop, gas, ignition, save, and timing are the same. Similarly, the position of the words "Horizontal and Vertical Burner - Vertical" is also the reset button, and clicking it will reset to the boot interface.
4. T1 is the first flame time, T2 is the second flame time, T3 is the flameless burning time to spread to the fixture and the bottom of the shop, and the corresponding style position at the lower end of the companion can be clicked to select the corresponding situation.
5. ta - flame application time. Intermittent time - the time it takes for the flame to return to the next auto-advance;
Number of cycles – i.e. the number of burns.
1) Clear Historical Data – Clicking Clear will clear all saved historical data.
2) The motor returns to turn off the gas - there are two options of "yes" and "no" in the switching state, select "yes", the system will turn off the gas at the same time as the motor returns to the spike bureau, and the gas should be turned on again in the operation interface before the next start, and the Bunsen lamp will be bright when you select "No".
3) Automatic ignition when the motor starts - there are two options of "yes" and "no" in the switching state. Only when the motor returns to turn off the gas and selects Yes, you can select "Yes", select automatic ignition, then the system will automatically turn on the gas ignition after pressing the start button three times and then automatically start the Bunsen burner.
-
50W UL94 horizontal and vertical fire test:
Go to the [500w Horizontal and Vertical Burning] screen.
1) Pull out the operating handle to make the burner (Bunsen burner.
Outside the seat, place the burner tube vertically, select the required combustion power (50W), and place the flow meter first.
Increase the gas flow slightly.
2) Press the [Burn Start] button to light the flame. Adjust the flowmeter knob, rotate the air conditioning device on the upper part of the Bunsen burner to adjust the height of the outer flame, adjust the adjustment knob of the air inlet under the Bunsen burner, adjust the height of the inner flame of the flame (20mm), and the two air intake knobs are adjusted by cooperating with each other to obtain the flame required by the standard. Flame regulation and temperature rise measurement are described in [Temperature Rise Measurement].
3) Set the parameter values of each time and times in [500W horizontal and vertical combustion], and install the samples according to the standard requirements.
4) Press the [Multiple Burn, Single Combustion] selection button to switch to single combustion selection, which is the UL94 50W horizontal and vertical combustion test.
5) Press the [Combustion Start] on the remote control or touch screen, the burner will be automatically ignited, and when the flame burns stably to the time required by the standard, adjust the distance between the burner angle and the sample according to the standard requirements, and close the window door. Push the burner to the limit position (burning position) through the handle, and press the [Burning Time Timer Start] button at the same time to start the timer. Burn time to extinguish the flame.
-
The vertical combustion test method is suitable for the determination of flame retardant properties of clothing fabrics, decorative fabrics, tent fabrics, etc. with flame retardant requirements.
-
The horizontal and vertical combustion tester is to simulate the early situation of fire in the surrounding environment of electronic and electrical products, and is used to evaluate the fire hazard of simulation technology, and ensure that the fire around the product does not cause combustion spread. The test equipment is suitable for components, parts and components of electrical and electronic household appliances, such as: insulating shells, switch panels, printed circuit boards and insulating materials of household appliances.
This model UL94 horizontal and vertical combustion tester is designed to test the flammability of equipment and appliance parts and materials, and many end-use test indicators such as flammability, combustion rate, flame spread, combustion strength and flame retardant performance of products can be tested. Learn more - Standard) Group.
Standards: 1. Horizontal Combustion Test: UL94HB (ASTM D 635, IEC 60695-11-10, IEC 60707, ISO 1210).
2. Vertical flame test: UL 94 V-0, V-1, or V-2 (ASTM D 3801, IEC 60695-11-10, IEC 60707, ISO 1210).
W (125mm) Vertical Flame Test: 5VA or 5VB (ASTM D 5048, IEC 60695-11-20, IEC 60707, ISO 9772) 500 W (125mm).
4. Vertical flame test of thin materials: VTM-0, VTM-1, or VTM-2 (ASTM D 4804, ISO 9773).
5. Horizontal foam material burning test: HF-1, HF-2 or HBF (ASTM D 4986, ISO 9772).
6. Burner :(ASTM D 5025, ASTM D 5207, ISO 10093, ISO 103351).
-
The horizontal and vertical combustion tester developed by the laboratory of Zhongchuang Alliance is to simulate the early situation of fire in the surrounding environment of electronic and electrical products, and use simulation technology to evaluate the fire hazard. The test equipment is suitable for electrical and electronic products, household appliances, components, parts and components, such as: household appliances insulating shells, switch panels, printed plates and insulating materials.
-
The series of horizontal and vertical combustion test equipment products are based on UL94, GB2408-2008, GB T5169-14 (IEC60695-11-2) "1KW Nominal Premixed Flame Equipment, Validation Test Methods and Guidelines", IEC60695-11-3 (5VA, 5VB grade material, flame power: 500W), IEC60695-11-4 (V-1 grade material, flame power: 50W), GB T5169-16 ( IEC60695-11-10) "50W Flame Test Method", GB T5169-17 (IEC60695-11-20) "500W Flame Test Method", etc., designed and manufactured to simulate the impact of flames in the early stage of ignition around electrical and electronic products, so as to assess the degree of fire risk through simulation technology.
It is mainly used for plastic and other non-metallic material parts samples, solid materials and ISO845 method determination of apparent density not less than 250kgm'The relative combustion characteristics of foam plastics are tested for horizontal and vertical flammability.
UL 94-HB: The lowest flammability rating in the UL94 standard. It is required that the burning rate is less than 40 mm per minute for samples thick from 3 to 13 mm; Samples less than 3 mm thick with a burning rate of less than 70 mm per minute; Or extinguish in front of a 100mm sign.
UL 94-V2: The flame is extinguished within 30 seconds after two 10-second burn tests on the sample. It can ignite cotton wool up to 30 cm.
UL 94-V1: The flame is extinguished within 30 seconds after two 10-second burn tests on the sample. Do not ignite cotton wool below 30cm.
UL 94-V0: The flame is extinguished within 10 seconds after two 10-second burn tests on the sample. No incendiary material can be dropped.
-
1. Plastic horizontal and vertical combustion testing machineComply with standards: GB T2408, GB13488, ISO1210, UL94 V-0 V-1 V-2, ASTM D635 4804 3801, IEC707, IEC 60695-11-10 standards.
2. Foam plastic horizontal combustion testing machineStandards: GB T8332-2008, ISO9772:2001.
3. Horizontal + vertical combustion testing machineCompliant Standards: VW-1 and CSA FT-1 compliant.
4. Vertical combustion testing machine for vertical pipesConform to the standard: meet the UL1666 <> test standard.
5. Vertical combustion testing machine for a single wire and cableStandards: GB:2004, IEC60332-1-2:2004, IEC60332-1-3:2004.
Comply with GB:2004, IEC60332-2-2:2004.
-
There are three kinds of horizontal and vertical combustion testing machines produced by Zhongnuo Instruments: 1. ZY6017 plastic horizontal and vertical combustion testing machine, in line with the standards: GB T2408, GB13488, ISO1210, UL94 V-0 V-1 V-2, ASTM D635, IEC707, IEC foam horizontal combustion testing machine, in line with the standard:
It conforms to GB T8332-2008 "Foam Plastic Combustion Performance Test Method - Horizontal Combustion Method" and ISO9772:2001 test standard.
3. ZY6017B-VB Rigid Foam Horizontal Combustion Tester, in line with the standard: in line with GB T8333-2008 "Rigid Foam Combustion Performance Test Method - Vertical Combustion Method" and ASTM D3014-04
<><1. The hazards in the safety standards include the following four types: electrical damage, mechanical and physical damage, low-voltage and high-energy injury, and flammable prevention and control, which are the most basic safety standards in various product safety regulations, among which for general electronic products, electrical damage is the most damaging to users. >>>More
There are generally two kinds of electronic formaldehyde testers on the market. >>>More
First of all, when you use the cable fault tester, you must understand whether its terminal port is normal, such as whether there is a rust problem, whether there is an aging problem, if these problems occur, you also need to replace the wire in time, before you can continue to use it, even if you know what the specific experience is, it will affect the entire detection effect after it is connected, and even say that the detected data is very inaccurate, then it will only bring more trouble to yourself. Therefore, when you need to understand its terminal before wiring, see if it is aging or rusty, and if so, you need to replace it in time to ensure the specific detection effect. >>>More
The test methods of transformer winding deformation mainly include low-voltage pulse analysis method, frequency response analysis method, impedance analysis method, water wave analysis method and ultrasonic analysis method. >>>More
If the internal resistance of the battery is higher than 50% or higher of the 100% reference value, the battery should be replaced. It is recommended to contact the manufacturer if the internal resistance is 25% to 43% above the benchmark value. At levels above 43%, it is load tested. >>>More